Digox 602 dac

導電度脫氣裝置

Degassed Acid Conductivity

廠牌:Dr. Thiedig
供應商:聯宙科技股份有限公司
聯絡電話:02-22320556
定價:與我們聯絡

快速詢價

9612

行動條碼QR Code
導電度脫氣裝置-行動條碼QR Code

The conductivity in water-steam cycle in power plants is an important measurement.

It must be distinguished between:

- Specifi c conductivity
which records the sum of all charge carriers and is mainly caused by enriched alkalising agents.

- Acid conductivity
In cation fi lter, the H+ from the exchanged cations combine with OH- from alkalizing agents to water. The remaining conductivity is determined by the autoprotolysis of the pure water plus the impurities in the form of anions, i.e. also CO32-.

To operate a steam turbine, the acid conductivity must not exceed a threshold of typically 0.2 μS/cm.

The following causes for increased conductivity are possible:
•unclean piping system, high corrosion conditions
•cooling water leakage in the condenser
•atmospheric air-in leakage with CO2-impact
•organic substances in the boiler feed water - CO2- impact after heating

- Degassed Cation conductivity
For the shortest possible start-up phase, the acid conductivity must be measured without the infl uence of the conductivity caused by dissolved CO2. Thus the threshold for operation of the turbine is reached faster. The dissolved CO2 rises up the conductivity, but does not harm the turbine.

Therefore, it is necessary to remove the carbonic acid from the sample and to measure the conductivity again (degassed acid conductivity). Thus, on one hand, the startup phase can be signifi cantly reduced. On the other hand, the system can be monitored for organic substances and atmospheric air ingress.

With the Digox 602 dac you have a universal measuring instrument at your disposal for this kind of tasks. In the compact design, you have fi rst a double conductivity measuring including a cation fi lter, automatic venting and pH calculation according to VGB-S006, then a separating operable degassing unit with conductivity measurings before and after degassing. Depending on the effi ciency, the degassing can be calculated to 100 %.

ADVANTAGES:
•Degassing and measurement of all conductivities at the same, not elevated sample temperature
•No heating up, therefore no gas emissions of other volatile acids
•No inert gas required, air-conditioning by means of air treatment
•High gain of degassed carbonic acid
•Very short response times t90 < 90 s for degassing unit
•Regenerative operating chemicals for cation exchanger
•Very low power consumption < 60VA
•Available as retrofit option for existing measurement of the cation conductivity: Digox dac basic
•Simple flow adjustment and stabilization with built-in flow stabilizer
•Improved efficiency of the degassing >90%, with switchable calculation to 100 %
•Highly accurate measurements of temperature and conductivity

The analyser Digox 602 dac ensures very short start-up times of the power plant and a simple, safe operation.

您也可能對以下商品有興趣...

產品關鍵字:

大台北尖端儀器有限公司

最新廠商動態!

WHATMAN代理美國過濾耗材知名大廠;Cytiva代理WHATMAN耗材在2020年;代理濾膜;濾紙;空氣採樣濾紙;
BROOKS質量流量控制器;PC540;BROOKS代理;質量流量控制器;BROOKS 5850E;
OMEGA ENGINEERING ;OMEGA 代理商;熱電偶接頭;熱電偶導線;熱電偶公母接頭;熱電偶感溫線;
Membrance Solution (USA) 0.22um/ 0.45um PES/ PVDF 針桶過濾膜特價供應
Membrance Solution (USA) 0.22um/ 0.45um 新款 PES 濾膜快速超低殘留型濾杯特價供應

更多

最新儀器!

力明儀器有限公司

Copyright © 2006-2017 tw17 All rights reserved.
Design: 台灣儀器網

台灣儀器網Tw17 行動條碼QR Code